Automatic Blood Pressure
Monitor & Printer
The World Hypertension
League recommends regular blood
pressure monitoring with clinically
validated devices.
Instruction manual
Gebrauchsanweisung
Gebruiksaanwijzing
page
1
Mode dꢀemploi
page 67
pagina 89
pagina 111
Seite 23
Istruzioni per lꢀuso
Instrucciones de uso
pagina 45
IM-HEM-759-E2-02-10/2011
1628532-8F
705CP-II English
Regular blood pressure monitoring is essential for the
prevention, the control and the management of
hypertension. It also helps to support the doctors' work.
The World Hypertension League is a world-wide
association of experts specialising in hypertension. The
World Hypertension League recommends regular blood
pressure monitoring by doctors and patients with
clinically validated devices.
The World Hypertension League
recommends regular
blood pressure monitoring by
doctors and patients with clinically
validated devices
As the worlds leading manufacturer OMRON provides a
range of products for regular blood pressure monitoring.
• Important: Blood pressure measurement is not
suitable in cases of serious arteriosclerosis
(hardening of the arteries).
Important
instructions to
help you obtain
meaningful
readings
• The pulse display is not suitable for monitoring the
frequency of cardiac pacemakers.
• If you suffer from disorders of heart rhythm, known
as arrythmia, you should only use this blood pressure
monitor in consultation with your doctor. In certain
cases the oscillometric measuring method can
produce incorrect readings.
• Pregnant women should only measure their own
blood pressure in consultation with their doctor, since
the readings may be changed by pregnancy.
• You should avoid eating, drinking (alcohol), smoking
and sports before measuring your blood pressure, as
this could affect your blood pressure level.
• Donꢁt move, donꢁt speak while measurement is being
taken.
• Make yourself comfortable in a chair and relax before
each reading.
• We recommend that you monitor your blood pressure
twice a day, in the morning after getting up and in the
evening after work, or as advised by your doctor.
• Please remember: Self-measurement is not the
same as medical treatment !
You should never change the dose of medicines
prescribed by your doctor.
2
705CP-II English
OMRON 705CP-II is a compact, fully automatic blood
pressure monitor, operating on the oscillometric
principle. It measures your blood pressure and pulse
rate simply and quickly. It contains an intelligent “fuzzy
logic” system for controlled inflation,
Your blood
pressure
monitor:
known as “Intellisense”. This is an
the benefits
advanced method of oscillometric
measurement and does not require pressure pre-
setting or re-inflation. OMRON is the worldꢁs leading
manufacturer of blood pressure monitors with the goal
to always fulfill the need for reliable regular monitoring.
Set Button
Heart Symbol
Adjust Button
AC Adapter Jack
USB Connector
Systolic Blood
Pressure
Diastolic Blood
Pressure
ON/OFF (O/I) button
START ( ) button
Air Jack
Pulse Rate
Date/Time Display
Memory recall Button
Cuff (Applicable cuff for Adults 22 to 32 cm.)
Air Plug (Connects to Air Jack)
Green Mark
Air Tube
3
705CP-II English
Slide the battery cover off in the direction of the
ꢀarrow.
Battery
installation /
replacement
Install or replace 4 “LR6” batteries so the +
ꢁ(positive) and – (negative) polarities match the
polarities of the battery compartment as indicated.
Replace the battery cover.
ꢂ
NOTES :
If the Low Battery Indicator (
) appears on the
display, replace all four batteries. Long-life alkaline
batteries are recommended.
Remove the batteries if the monitor will not be
used for an extended period of time.
Replace batteries within 30 seconds. If the
batteries remain removed longer than that time,
the set date and time will be deleted.
4
705CP-II English
Install the batteries and press the ON/OFF button.
ꢀThe display will show 0:00
How to set time
and date
Press and hold the SET button.
ꢁThe year digits (2002) will flash.
Press the ADJUST button to advance the digit(s)
ꢂone at a time.
If you hold down the ADJUST button, the digits will
advance rapidly.
DATE/TIME SET
SET
ADJUST
Press SETꢂbutton when desired number is on the
ꢃdisplay to lock setting.
Repeat Step 3-4 for month and date.
Repeat Step 3-4 for hour and minutes.
ꢄ
ꢅ
5
705CP-II English
• You can wrap the cuff either on your right or left arm.
How to apply
the cuff
(when wrapping on your left arm)
Insert air tube to air jack (on the left side of the
ꢀdivice). The cuff must be airless.
Remove tight-fitting clothing from your upper arm.
ꢁ
Sit in a chair with your feet flat on the floor and
ꢂplace your arm on a table so that the cuff is at the
same level as your heart.
Put your arm through the cuff loop. The bottom of
ꢃthe cuff should be approximately 1 to 2 cm above
the elbow. The green marker on the cuff should lie
over the brachial artery on the inside of the arm.
The tube should run down centre of arm
approximately even with middle finger.
Bottom Edge 1 to 2 cm
Cuff Tube
Green Mark
6
705CP-II English
Pull the cuff so that the top and bottom edges are
ꢄtightened evenly around your arm.
When the cuff is positioned correctly, close the
ꢅvelcro fastener FIRMLY.
Make certain the cuff fits snugly around your arm.
ꢆThe cuff should make good contact with your skin.
You should be able to fit your index finger
between the cuff and your arm easily, so you
can pull the cuff off and on.
Relax your arm and turn your palm upward.
Be sure there are no kinks in the air tubing.
ꢇ
Right hand measurement
7
705CP-II English
Press the (O/I) button.
ꢀ
How to take a
reading
a) All display symbols appear for approximately
one second.
b) When the monitor becomes ready to measure,
the (
) symbol appears on the display.
Press the (START ) button and remain still.
ꢁ
As the cuff begins to inflate, the monitor
automatically determines your ideal inflation level.
Because this monitor detects the pulse even
during inflation, do not move your arm but remain
still until the entire measurement completes.
• If you want to stop the inflation or measurement,
press the (O/I) button. The monitor will stop
inflating and start deflating rapidly, then the
monitor will turn off.
Inflation
Inflation stops automatically and measurement is
ꢂstarted.
As the cuff slowly deflates, decreasing numbers
appear on the display and the (
) symbol
Deflation
flashes at every heartbeat. In rare circumstances,
a higher inflation may be necessary. In those
cases, the monitor reinflates the cuff up to 30
mmHg higher than initial inflation and restarts the
measurement.
8
705CP-II English
When the measurement is complete, the cuff
ꢃcompletely deflates and your blood pressure and
pulse rate are displayed.
NOTE :
The monitor automatically stores blood pressure
and pulse rate into the memory.
• When 28 sets of readings are stored in memory,
the oldest set will be deleted to store a new set.
Press the (O/I) button to turn the monitor off.
ꢄ
NOTE :
If you forget to turn the monitor off, it will
automatically shut itself off after five minutes.
Instructions for special conditions:
• If your systolic pressure is known to be more than
220 mmHg, press and hold the (START ) button
until the monitor inflates 30 to 40 mmHg higher
than your suspected systolic pressure.
START
NOTE :
Continue to press button
Do not apply more pressure than necessary.
The monitor will not inflate above 300 mmHg.
9
705CP-II English
This monitor has a memory capable of storing 28
How to use the
memory
sets of readings. Every time you complete the
measurement, the monitor automatically stores blood
pressure and pulse rate.
function
• Turn on the monitor by pressing the (O/I) button.
• Press the (
) button to recall stored readings
while the ( ) symbol is displayed.
Date and Time of stored readings
will be alternately displayed.
• To delete all the stored readings, press the (
and the (START ) button simultaneously.
) button
NOTE :
You can not delete the stored readings partially.
START
Press these two buttons
at the same time.
10
705CP-II English
• Before measuring, stretch your back and sit up
straight, breath normally 5-6 times, relax your
shoulders, arms and entire body.
Hints on taking
blood pressure
readings
• Slightly bend and support your elbow (e.g. on a
table).
• Do not move, talk or touch the device during
measurement.
• Stress influences your blood pressure, so try to
relax before taking a measurement.
• To compare results, measurements must always
be taken from the same arm.
• Do not place the cuff over thick clothes and do not
roll up your sleeve if it is too tight.
• Align the cuff with the height of your heart (breast
height). If necessary use a pillow or cushion.
• Coldness raises your blood pressure. It is
recommended to measure at room temperature
(+20°C).
• Relax before measurement. If necessary go to the
bathroom before measurement.
• Do not measure directly after bathing.
• If you wish to interrupt measurement prematurely,
press the (O/I) button. Inflation is stopped and the
air is automatically released from the cuff.
• Blood pressure measurements taken in a moving
vehicle may be influenced by vibration.
• If an error has occurred during measurement
displayed. Please refer to the next chapter.
E
is
• The monitor should be switched off after
measurement to avoid running down the batteries
unnecessarily. If you forget to do so, the monitor
switches itself off automatically after 5 minutes.
11
705CP-II English
Failure, causes and rectification
Failure and their possible causes
Rectification
Incorrect measurement is indicated by
E
1. A correct reading could not be obtained
because measurement was disturbed by
movement of the body.
1. Repeat the measurement keeping
perfectly still. Do not move your arm and
do not speak.
2. The cuff is not fitted correctly.
2. Check that the cuff is correctly fitted.
Then repeat the measurement.
3. Your clothing has constricted your
blood flow.
3. Remove the item of clothing which
caused the constriction.
4. There is still air in the cuff when the
monitor is switched off.
4. The unit may be defective. Please have
it examined and, if necessary, repaired
by OMRON.
The (START ) button was pressed
Wait for the ( ) symbol indicating
before the (
) symbol was displayed.
readiness before you press the (START
button.
)
The blood pressure values displayed are
extremely low or high, or they are
implausible.
Refer to the instructions and then repeat
the measurement.
The display does not light up when the
(START ) button is pressed.
1. The batteries are exhausted.
1. Check the batteries and, if necessary, fit
4 new batteries.
2. The +/- poles of the batteries have been
reversed.
2. Reinsert the batteries with the +/- poles
the right way round.
3. The battery contacts are dirty.
3. Clean the battery contacts with a dry
cloth.
The batteries are weak or exhausted.
Fit new batteries.
The cuff pressure does not rise
Check that the air tube is properly
connected to the monitor. Push the tube
connector firmly into the socket.
although the pump motor can be heard.
problem with mem. function
Contact your Omron retailer or distributor.
Er
12
705CP-II English
• For cleaning the monitor only use a soft, lightly
Maintenance
and storage
moistened cloth.
Do not use petroleum spirits, thinners or
similar solvents !
• Stains on the cuff can be carefully removed using
a moist cloth and soapy water.
Do not wash the cuff or make it wet !
• Keep your blood pressure monitor in the storage
case to protect it from dust and moisture.
• Protect your monitor against vibrations and do not
leave it in a place where temperatures are
extremely low (below -20°C) or extremely high
(above 60°C) or with high humidity (+ 85%).
• Do not use your monitor at very low temperatures
(below 10°C) or very high temperatures (above
40°C).
• Do not fold the cuff or air tube together too tightly
and do not crease them.
• Do not carry out repairs of any kind yourself.
If a defect occurs, consult the OMRON distributor
or Customer Services or seek the advice of your
surgical appliance stockist or pharmacist.
• The accuracy of this blood pressure monitor is
designed to last a long time.
• In order to increase the lifetime of the batteries
take them out when you store the device for a
longer period than 2 months.
• Electro-magnetic fields can influence the
measurement. Do not use a portable phone near
this unit.
13
705CP-II English
This device fulfils the provisions of the EC directive
93/42/EEC (Medical Device Directive). This blood
pressure monitor is designed according to the
European standard EN1060, Non-invasive sphygmo-
manometers Part 1: General requirements and Part
3: Supplementary requirements for electromechani-
cal blood pressure measuring systems. This OMRON
product is produced under the strict quality system
of OMRON HEALTHCARE CO. LTD., Japan. The Core
component for OMRON blood pressure monitors,
which is the Pressure Sensor, is produced in Japan.
Technical data
Model:
OMRON 705CP-II
Display:
LCD digital display
Blood Pressure
Measurement Range:
Accuracy:
Pressure: 0 mm Hg to 299 mm Hg
Pulse: 40 – 180/min.
Pressure: 3 mm Hg
Pulse: 5% of reading
Memory:
28 measurements
Inflation:
Fuzzy-Logic controlled by electric pump
Automatic pressure release valve
Capacitive pressure sensor
Oscillometric method
Deflation:
Pressure Detection:
Measurement Method:
Power Source:
4 Alkaline batteries 1.5V (Type LR6) or
AC/DC adapter (optional, 6V = 4W)
Capacity of new batteries is approx. 300 measurements
10°C to 40°C
Battery Life:
Operating Temperature:
Storage Temperature:
Humidity:
-20°C to 60°C
15 to 90% RH maximum
Storage humidity:
Console Weight:
Outer Dimensions:
Cuff Dimensions:
10 to 95% RH maximum
Approximately 380g without batteries
Approximately 115 (l) mm x 177 (w) mm x 71 (h) mm
Approximately 140 mm x 480 mm
(arm circumference 22 – 32 cm)
Medium cuff, instruction manual, storage case, battery
set
Accessories:
Made in China
NOTE: Subject to technical modification without prior notice
• Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
= Type B
14
705CP-II English
Calibration
OMRON spare
parts
The accuracy of this blood pressure monitor has
been carefully tested and is designed for a long
service life. It is generally recommended to have the
monitor inspected every two years to ensure correct
functioning and accuracy. Please consult your
authorised dealer or the OMRON Customer Service
at the address given on the packaging/attached
literature.
AC adapter
OMRON devices are especially designed for regular
blood pressure monitoring. OMRON therefore has
direct contact with medical specialists and takes
advice, on the design and features of its blood
pressure monitors.
Small cuff
Arm circumference 17 – 22 cm
Spare parts & accessories
Due to high company quality standards, OMRON
considers the main unit as a non-serviceable part
because of the necessity of proper calibration after
replacement of high-tech components.
Medium cuff
Arm circumference 22 – 32 cm
Extra large cuff
Arm circumference 32 – 42 cm
15
705CP-II English
Some useful information about blood pressure
What is Blood Pressure ?
Blood pressure is a measure of the force of blood flowing against the walls of the
arteries. Arterial blood pressure is constantly changing during the course of the heartꢁs
cycle. The highest pressure in the cycle is called the SYSTOLIC BLOOD PRESSURE;
the lowest is the DIASTOLIC BLOOD PRESSURE. Both pressure readings, the
SYSTOLIC and DIASTOLIC, are necessary to enable a doctor to evaluate the status of
a patientꢁs blood pressure.
Many factors such as physical activity, anxiety, or the time of day, can influence
your blood pressure. Blood pressure is typically low in the morning and increases from
afternoon to evening. It is lower in the summer and higher in the winter.
Blood pressure is measured in millimetres of mercury (mmHg) and measurements
are written with the systolic pressure before the diastolic e.g. A blood pressure written
as 140/90, is referred to as 140 over 90 mmHg.
High Blood Pressure
The discovery that you have high blood pressure is more often than not a chance
finding. People with high blood pressure usually feel well unless the blood pressure has
been high for some time, and complications have occurred. Such complications may
effect the heart, brain and other important organs. They may bring on ill health and
affect your enjoyment of life. It is because of the complications, rather than the high
blood pressure itself, that this condition is taken so seriously today.
High blood pressure is sometimes referred to as raised blood pressure, elevated blood
pressure, or hypertension. Most people with this condition require regular long-term
drug treatment, usually in the form of tablets. Some can lower their blood pressure
sufficiently by other means, such as changing diet and life-style.
People who have high blood pressure, and who also understand their condition, have
an advantage. They are more likely to attend for regular check-ups, to cooperate with
advised diets and changes in lifestyle, and to take medicines correctly. This
immediately leads to better health. It is hoped that, by a better understanding of high
blood pressure and how it is treated, you will take an active and informed part in your
health care, and so enjoy a better quality of life.
16
705CP-II English
What Causes High Blood Pressure ?
Blood pressure is only classed as high if it doesnꢁt go down when you rest. Permanently raised blood
pressure can be caused by several factors such as hardening of the arteries, smoking or drinking too
much alcohol. It can also be caused by cholesterol, a type of fat, building up inside your blood vessels
and making them narrower. Being under stress can add to the problem.
High blood pressure is also known to run in families hence if you find you have a high blood pressure it
is a good idea to test other closely related members of your family.
Fluctuation within a Day
(Example: 35-year-old male)
Blood pressure
(mmHg)
Breakfast
Dinner
Sleep
Lunch
Rest room
Systolic
blood
pressure
Diastolic
blood
pressure
Time
Brachial blood pressure data
Why is it a Good Thing To Measure Blood Pressure at Home ?
Having your blood pressure measured by a doctor can cause anxiety which is a cause itself of high
blood pressure. As a variety of conditions affect blood pressure, a single measurement may not be
sufficient for an accurate diagnosis.
The blood pressure measured first thing in the morning after getting up, before eating and while at rest,
is known as the fundamental blood pressure. In practice it is rather difficult to record the fundamental
blood pressure, but it is important to get readings in an environment that is as close as possible to this.
Systolic blood pressure
(mmHg)
Classification of Blood Pressure by the World Heath
Organization
Severe hypertension
The World Health Organization (WHO) and the International
180
Moderate hypertension
Society of Hypertention (ISH) developed the Blood Pressure
160
Classification shown in this figure.
Mild hypertension
140
This classification is based on the blood presure values
Normal systolic value
measured on people in a sitting position in outpatient
130
Normal blood pressure
departments of hospitals.
120
*There is no universally accepted definition of hypotension. However, those
Optimal blood
pressure
Diastolic
blood
having the systolic pressure below 100 mmHg are assumed as hypotensive.
pressure
(mmHg)
(target value)
90
100
110
80
85
According to the blood pressure classification by the WHO/ISH* (revised in 1999)
* International Society of Hypertension
17
705CP-II English
The OMRON printer can be conected to OMRON
blood pressure monitors with IT functions.
Introduction
Description
Printer Cover
Printer Cable Plug
Power/Operation
Light
Paper Feed
Print Graph
Print Current Data/Stop Printing
Print All Data
Slide the battery cover off in the direction of the
ꢀarrow.
Battery
installation /
replacement
Install or replace 4 “LR6” batteries so the +
ꢁ(positive) and – (negative) polarities match the
polarities of the battery compartment as indicated.
Replace the battery cover.Remove the batteries if
ꢂthe monitor will not be used for an extended
period of time.
18
705CP-II English
Remove printer cable plug from the printer unit.
ꢀ
How to connect
the printer unit
Connect printer cable plug cord to the main unit.
ꢁThe circle on the cable plug should face up.
Open printer cover in the direction indicated by
ꢀthe arrow symbol and remove paper roll from the
How to load the
printer paper
main unit.
Peel of the glued edge of the printing paper
ꢁincluded with the unit, and cut the first 10 cm (4
inch) with a pair of scissors.
Press theꢂFEED button while inserting the edge
ꢂof the printing paper into the paper feeding slot.
4in
When the edge of the paper protrudes sufficiently,
release the ꢂFEED button.
Close the plastic printer cover. Make certain that
ꢃthe paper sticks out from the top when the cover
is replaced.
Thermal sensitive paper is supplied with the unit.
ꢄUse only thermal paper of similar size (width
approx. 38mm) and roll diameter (not to exceed
approx. 26mm)! These rolls of paper can be
ordered from OMRON. Printer paper can be
ordered by your OMRON shop or distributor.
19
705CP-II English
OMRON IT blood pressure monitors have the ability
to print all the readings stored in memory.
How to use the
printer
NOTE: • Make sure printer-unit is connected to main
unit (see How to Connect Printer-Unit)
• Do not use printer-unit without loading
printer paper (see How to Load Printer
Paper). Improper use will decrease
operation life cycle of printer-unit.
• Main unit does NOT have to be ON to
operate printer-unit.
• Printer-unit will automatically shutꢂOFF
when printing is completed.
• If you need to stop printing during printing
process, press “DATA/STOP” button.
Press “DATA/STOP” to print measurement of
ꢀmost recent reading.
The numeric print-out will look like the
following example:
Month/Date
Time
7/9
6:36 AM
Systolic Reading
Diastolic Reading
Pulse Reading
SYS
111mmHg
67mmHg
77/min
DIA
PULSE
Press “ALLꢂDATA” button to print a numerical
ꢁlist and average of all stored readings ( maximum
depending on your blood pressure monitor).
The numerical print-out will list all stored readings
starting with the most recent reading to last stored
reading:
Systolic Reading
Column
Diastolic Reading
Column
Month(s)
3-6
D. T.
mmHg
SYS
/min
PUL
Pulse Reading
Column
DIA
Date
20 15
23 11
1 15
3 14
5 7
15 7
20 12
3 12
10 12
12 14
14 7
21 15
22 12
4 6
113
117
113
110
112
110
106
110
108
111
107
111
121
112
112
75
69
71
67
66
61
66
65
61
56
64
64
62
64
65
67
64
63
66
64
64
65
64
67
62
64
64
66
63
65
Oldest reading
Time/Hour
Most recent reading
Average of all stored
readings
*AV.
20
705CP-II English
• Month(s)
The month or months will be printed.
Example: If the blood pressure reading was stored
between months March-June, then the printout will
indicate “3-6”.
• Date
Only the dayꢁs date (1-31) will be printed.
Example: If the blood pressure reading was stored
on June 1st, then the printout will indicate 1 under
the “D” column.
• Time/Hour
Only the hour (in military time) will be printed.
Example: if the blood pressure reading was stored
at 3:00 PM, then
the printout will indicate 15 under the “T” column.
Note: If Time and Date was not initially set up
before taking blood pressure measurement (see
How to Set Time and Date) on printer-unit, the
numerical value will not appear on print-out for
month, time/hour and date.)
Press “GRAPH” button to print a graph
ꢂof all stored readings
Month(s)
3-6
50
mmHg
200
100
150
Oldest reading
If reading is lower than 50
mmHg or higher than 200
mmHg, will appear.
Systolic Reading
Diastolic Reading
Most recent reading
Average diastolic reading of
stored measurements
Average systolic reading of
stored measurements
21
705CP-II English
Troubleshooting
Error Indicator
Cause
Correction
Printer cable plug is not
connected properly or damaged
Connect correctly reference p.10
Contact your OMRON distributor.
Nothing is printed
Replace with new batteries
Take measurement then print
Batteries are worn
No readings stored
Printer paper is not installed
correctly
Install correctly
Power/Operation light is blinking,
stopped printing
Batteries are worn
Replace with new batteries
No printer paper installed
Printer paper jammed
Ran out of printer paper
Take paper jammed from the
printer and install again
No printer paper comes out
Purchase new printer paper
Specifications
Power Source:
Battery Life:
4 “AA” batteries
Approximately 300 uses
Paper Usage:
Approximately 200 uses for 1 roll
(Printer) Weight:
Approximately 6 oz. (170 g) (not including
batteries)
(Printer) Dimensions:
Approximately 4 4ꢃ5'' (l) x 2 4ꢃ5'' (w) x 2'' (h)
(123 mm x 72 mm x 50 mm)
NOTE: These specifications are subject to change without notice.
EU-representative:
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
Disposal of this product and used batteries should be carried out in
accordance with the national regulations for the disposal of electronic
products.
22
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
EU-representative
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
EC
REP
Production facility
Produktionsstätte
OMRON (DALIAN) CO., LTD.
Dalian, CHINA
Productiefaciliteit
Site de production
Oficina de producción
Stabilimento di produzione
Subsidiary
OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.
Niederlassung
Dochteronderneming
Succursale
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
John-Deere-Str. 81a, 68163 Mannheim, GERMANY
Empresa filial
Consociata
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE
Uniquement pour le marché français: OMRON Service Après Vente
Nº Vert 0 800 91 43 14
Made in China
Hergestellt in China
Geproduceerd in China
Fabriqué en Chine
Fabricado en China
Prodotto in Cina
|